Automatic Control:
- Fraction Collector
- Vacuum Control
- Condenser Temperature
- Reflux Ratio
- AET Temperature Calculation
- Distillation Column Heated Jacket Control
- Recording of Distillation Data
- Distillation Rate Control
- Ending of Distillation and Cool Down
Semi-Automatic Control:
- Heat Control for Boiling Flask
Operator Controlled:
- Measurement of Receiver Volume
- Creating Volume vs Temperature Distillation Curve
Boiling Flask/Heating Mantle:
- Stirring Heating Mantle
- Heated Mantle Top
- Stainless Steel Boiling Flask (optional glass flasks available)
- Quench Coil for Rapid Cooling of the Flask Contents
- PRT Temperature Probe to Monitor Boiling Flask Temperature
- Boiling Flask Sizes 1, 2, 3, 6, 12, 22, 30, 50, 100 liters (Other sizes available)
ASTM D2892 Distillation Column:
- According ASTM specifications
- Vertical viewing window for easy viewing
- Highly reflective vacuum jacket
- Heated Jacket to Prevent Heat Loss (Automatically Controlled)
- 15 Theoretical Plates (Other Efficiencies Available)
ASTM D2892 Distillation Column Sizes, Propak:
Diameter |
Packed Height |
Sample Range |
70 mm |
91 cm |
15-60 liters |
50 mm |
80 cm |
6-24 liters |
25 mm |
57 cm |
1.2-4.8 liters |
Other D2892 Distillation Columns:
We also offer distillation columns packed with mesh packing (Goodloe) and real plate distillation columns.